Analysis and Solution of Common Problems in Environmental Testing Instruments


1. In the humidity and humidity test, the actual humidity will reach 100% or the actual humidity and the target humidity will differ greatly. The value is much lower. The former phenomenon may be due to dry gauze on the wet ball sensor. It is necessary to check whether the water tank in the wet ball sensor is short of water. The water level in the water tank is automatically controlled by a water level controller. The pipeline pump checks whether the water supply system of the water level controller is normal and whether the water level controller is working properly. Another possibility is that the wet ball gauze may cause the gauze to become hard due to the long use time or the purity of the water supply, so that the gauze cannot absorb the moisture and dry, and the above phenomenon can be eliminated simply by replacing or cleaning the gauze. The latter phenomenon is mainly because the humidification system is not working. Check the water supply system of the humidification system. Check if there is a certain amount of water in the water supply system. Control whether the water level of the humidifying boiler water level is normal and whether the water level in the humidifying boiler is normal. If all of the above are normal for the magnetic pump, it is necessary to check the electrical control system. This requires professional maintenance personnel to perform maintenance.

2. In the high temperature test, if the temperature does not reach the test temperature value, the electrical system can be checked and the faults can be eliminated one by one. If the temperature rises very slowly, it is necessary to look at the wind circulation system and see if the adjustment baffle of the wind circulation is normally open. Conversely, check whether the motor operation of the wind circulation is normal. Sewage pumps, such as temperature overshoots, then need to set PID settings. If the temperature rises directly, over-temperature protection, then the controller fails and the control instrument must be replaced.

3. If the low temperature does not reach the test target, then you must observe the temperature change, whether the temperature drop is very slow, or whether the temperature will rise after the temperature reaches a certain value. The former must check whether it will be before the low temperature test. After the drying in the studio, the ambient temperature of the equipment such as the constant temperature and humidity box is kept dry, and the test sample is put into the working chamber for further testing. The test sample in the working chamber is placed excessively, and the chemical pump cannot make the wind in the working chamber dry. Full circulation, after eliminating the above reasons, we must consider whether it is a fault in the refrigeration system, so we must ask the factory's professionals for maintenance. The latter phenomenon is caused by the poor use environment of the equipment. The ambient temperature of the equipment and the location of the equipment (the distance between the cabinet and the wall) must meet the requirements (in the operating instructions of the equipment).

4. If the constant temperature and humidity box suddenly fails during the test operation, the control instrument displays a corresponding fault display prompt and a voice alarm. The operator can quickly check which type of fault belongs to in the troubleshooting section of the operation of the equipment, and the professional can quickly eliminate the fault to ensure the normal operation of the test. Other environmental test equipment will have other phenomena in use, it must be specific phenomena, specific analysis and exclusion.

Constant temperature is constant test equipment and other environmental testing equipment must be regularly maintained, the refrigeration system condenser regularly cleaned up, for moving parts should be according to the instructions oil lubrication, electrical control system regular maintenance inspections, etc., these tasks are essential.

EVA Raincoat

EVA Raincoat, Womens EVA Raincoat, Mens EVA Raincoat

Zhejiang CC Industrial And Trading Co., Ltd , https://www.ccraincoats.com